Independent Dimensional Uncertainty Budget

How do these independent measurement contributions combine?

Combine repeatability, calibration, resolution and one bounded dimensional contribution into standard and expanded uncertainty.

Open the calculatorRuns in your browser. No account needed.

How this tool works

Inputs, units and where to find them

InputUnitTypical source
Reported resultchoiceChoose a sample mean or one individual observation.
Sample standard deviationmmSample standard deviation from independent repeated measurements under the relevant conditions.
Independent observations in the sampleobservationsObservations used to estimate the standard deviation; also the averaging count in mean mode.
Calibration expanded uncertaintymmExpanded uncertainty from the calibration certificate in millimetres at the relevant point.
Calibration coverage factor×Coverage factor stated on that certificate; do not assume every certificate uses 2.
Digital resolution incrementmmSmallest digital increment, modelled as independent rounding over ±half an increment.
Other rectangular half-widthmmHalf-width of one independent rectangular contribution, already converted to equivalent dimensional error in millimetres.
Output coverage factor×Coverage factor justified by the uncertainty procedure and effective degrees of freedom.

What each input means

Open an input to see what it is, its units, the worked example and the common mistakes, with a small diagram.

What this means: Reported result
Illustration: Reported resultThe options for Reported result, with the one currently selected highlighted.Mean of the entered observationsOne observationExample: Mean of the entered obser…Reported result
What it is
Whether the reported measurement result will be the average of the repeated readings or a single reading. It decides whether repeatability is divided by the square root of the number of observations.
Units and format
Choose one of 2 options from the list: Mean of the entered observations; One observation.
Example
The form opens with “Mean of the entered observations” selected as a worked example. Choose the option that matches your case.
Effect on the result
The mean option shrinks the repeatability contribution by the square root of the observation count.
Get it right
Choose one observation if, in practice, a part is measured once; the mean option then would understate the uncertainty.
What this means: Sample standard deviation
Illustration: Sample standard deviationA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Sample standard deviation.upperlowersExample: 0.01 mmSample standard deviation
What it is
The spread of repeated measurements of the same feature, taken independently under the conditions that apply in use, calculated as a sample standard deviation.
Units and format
Enter a number in mm: millimetres. The form accepts 0 mm to 1000000 mm. Required: the tool does not calculate until every input is valid.
Example
The form opens with 0.01 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Its contribution is the standard deviation itself, or divided by √n for a mean.
Get it right
Repeat the full measurement each time (re-fixture, re-zero); repeated readings without resetting understate the spread.
What this means: Independent observations in the sample
Illustration: Independent observations in the sampleA row of identical items being counted; Independent observations in the sample is how many there are.nExample: 10 observationsIndependent observations in the sample
What it is
How many independent repeated measurements were used to calculate the standard deviation. In mean mode it is also how many readings are averaged for the result.
Units and format
Enter a number in observations: observations. The form accepts 2 observations to 100000 observations. Use a whole number. Required: the tool does not calculate until every input is valid.
Example
The form opens with 10 observations as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
In mean mode a larger count reduces the repeatability contribution; in single mode it has no effect.
Get it right
Count only independent repeats; several readings from one setup do not each count.
What this means: Calibration expanded uncertainty
Illustration: Calibration expanded uncertaintyA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Calibration expanded uncertainty.upperlowerUExample: 0.004 mmCalibration expanded uncertainty
What it is
The expanded uncertainty stated on the calibration certificate of the instrument or standard used, at the size you are measuring.
Units and format
Enter a number in mm: millimetres. The form accepts 0 mm to 1000000 mm. Required: the tool does not calculate until every input is valid.
Example
The form opens with 0.004 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
It is divided by the certificate coverage factor to give the calibration standard uncertainty.
Get it right
Convert certificate values to millimetres; certificates often quote µm, and 4 µm is 0.004 mm.
What this means: Calibration coverage factor
Illustration: Calibration coverage factorTwo bars: a base value and the same value multiplied by Calibration coverage factor, the factor.base valuewith factorkExample: 2 ×Calibration coverage factor
What it is
The coverage factor k that the calibration certificate used to turn its standard uncertainty into the expanded uncertainty you entered.
Units and format
Enter a number in ×: a multiplier (type 1.5 for 1.5 times). The form accepts 0.001 × to 100 ×. Required: the tool does not calculate until every input is valid.
Example
The form opens with 2 × as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
A larger certificate factor makes the calibration contribution smaller.
Get it right
Read k from the same certificate; do not assume a value if the certificate states another.
What this means: Digital resolution increment
Illustration: Digital resolution incrementA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Digital resolution increment.upperlowerrExample: 0.01 mmDigital resolution increment
What it is
The smallest step the instrument display can show, for example the last digit of a digital caliper or micrometer. Rounding to this step adds uncertainty.
Units and format
Enter a number in mm: millimetres. The form accepts 0 mm to 1000000 mm. Required: the tool does not calculate until every input is valid.
Example
The form opens with 0.01 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
The resolution contribution is the step divided by √12.
Get it right
Enter the full display step, not half of it; the tool converts it with √12 as a rectangular distribution.
What this means: Other rectangular half-width
Illustration: Other rectangular half-widthA nominal size line with an upper and a lower limit around it; parts inside the band are acceptable. The figure illustrates Other rectangular half-width.upperloweraExample: 0.003 mmOther rectangular half-width
What it is
Half the width of one other independent error source with known limits, such as a temperature effect, already converted into an equivalent dimensional error.
Units and format
Enter a number in mm: millimetres. The form accepts 0 mm to 1000000 mm. Required: the tool does not calculate until every input is valid.
Example
The form opens with 0.003 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Its standard uncertainty is the half-width divided by √3.
Get it right
Enter the half-width (±a), not the full range; enter 0 if there is no such contribution.
What this means: Output coverage factor
Illustration: Output coverage factorTwo bars: a base value and the same value multiplied by Output coverage factor, the factor.base valuewith factorkExample: 2 ×Output coverage factor
What it is
The coverage factor k used to expand the combined standard uncertainty into the reported expanded uncertainty, chosen by your uncertainty procedure.
Units and format
Enter a number in ×: a multiplier (type 1.5 for 1.5 times). The form accepts 0.001 × to 100 ×. Required: the tool does not calculate until every input is valid.
Example
The form opens with 2 × as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
Effect on the result
Expanded uncertainty is directly proportional to this factor.
Get it right
Justify k from your procedure and the effective degrees of freedom; it is a separate choice from the certificate factor.

Method and formulas

  1. Repeatability u = s/√n for a mean or s for one observation. Calibration u = certificate expanded uncertainty divided by its coverage factor.
  2. Resolution u = increment/√12. Other rectangular u = half-width/√3. Combined u is the square root of the sum of the four squared contributions. Expanded U = k × combined u.
  3. Example: s = 0.01 mm, n = 10, calibration U = 0.004 mm at k = 2, resolution 0.01 mm and half-width 0.003 mm gives combined u ≈ 0.005033223 mm and U ≈ 0.010066446 mm at output k = 2.

Limitations

  • All contributions must be in millimetres with unit sensitivity coefficients. Covariances and general propagation coefficients are not modelled.
  • Do not count resolution twice if another contribution already includes it. Resolution is not reduced by averaging here.
  • k = 2 does not automatically establish 95% coverage. Small samples, distributions, bias corrections and degrees of freedom need review.
  • An all-zero budget is rejected. No conformity decision or accredited certificate is issued.
  • Default values are worked examples, not recommended settings or material properties.

Typical uses

  • measurement budget
  • repeatability
  • coverage factor

Who it is for

  • Metrology teams
  • Quality engineers
  • Inspectors

What you get

  • Combined standard uncertainty
  • Expanded uncertainty
  • Component standard uncertainties and variance shares

Interpreting the result

  • Document the measurement model and independence evidence before using the combined result.
  • Variance shares help prioritise improvement. Record the quantity, unit, coverage factor and uncertainty together.

References

  • NIST TN 1297: Law of propagation of uncertainty — https://www.nist.gov/pml/nist-technical-note-1297/nist-tn-1297-appendix-law-propagation-uncertainty
  • NIST TN 1297: Reporting uncertainty — https://www.nist.gov/pml/nist-technical-note-1297/nist-tn-1297-7-reporting-uncertainty
  • Machinery’s Handbook, 29th edition (Industrial Press, 2012), printed pp. 709–738. Topic reference; implementation and examples are original, and no source tables are reproduced.