Guard-Banded Acceptance Decision (Uncertainty Budget, Accept, Reject or Indeterminate)
Given my gauge’s measurement uncertainty, can I accept this part from its reading, must I reject it, or is it too close to the limit to decide?
Acceptance decision for a measured part that allows for measurement uncertainty: the uncertainty budget from calibration, resolution, repeatability, Abbe error and temperature, the expanded uncertainty and test uncertainty ratio, and the guard-banded acceptance zone (as the measurement uncertainty budget tool), then the deviation of the reading from nominal compared with the guard-banded and the tolerance limits to accept, reject or flag it as indeterminate.
Open the calculatorRuns in your browser. No account needed.
How this tool works
Inputs, units and where to find them
| Input | Unit | Typical source |
|---|---|---|
| Nominal size | mm | Nominal size of the feature (mid-tolerance). |
| Tolerance band (total, symmetric) | mm | Total tolerance band, placed symmetrically about the nominal. |
| Measured size | mm | Size measured on the part. |
| Uncertainty contributors | ||
| Calibration uncertainty of the gauge (expanded, k = 2) | µm | Expanded uncertainty from the gauge’s calibration certificate. |
| Resolution of the gauge | µm | Smallest step the gauge displays. |
| Repeatability (standard deviation of repeat readings) | µm | Standard deviation of repeated readings on the same part. |
| Abbe offset | mm | Distance between the measuring line and the scale line (Abbe offset). |
| Tilt of the moving element | arcsec | Possible tilt of the moving jaw or slide. |
| Length measured | mm | Length being measured. |
| Part expansion coefficient | µm/m·K | Expansion coefficient of the part. |
| Uncertainty of the temperature difference | K | Uncertainty in the temperature difference between part and gauge. |
| Coverage factor | Coverage factor for the expanded uncertainty (2 for about 95%). | |
| Least test uncertainty ratio | Least ratio of tolerance to measurement uncertainty you accept. | |
What each input means
Open an input to see what it is, its units, the worked example and the common mistakes, with a small diagram.
What this means: Nominal size
- What it is
- The nominal size of the measured feature, taken here as the middle of the tolerance band.
- Units and format
- Enter a number in mm: millimetres. The form accepts -100000 mm to 100000 mm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 25 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- If the drawing tolerance is unequal, enter the mid-tolerance size, not the basic size.
What this means: Tolerance band (total, symmetric)
- What it is
- The total width of the tolerance band, from the lower to the upper limit, placed equally either side of the nominal.
- Units and format
- Enter a number in mm: millimetres. The form accepts 0.0001 mm to 100 mm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 0.05 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- A wider band widens the acceptance zone and raises the test uncertainty ratio.
- Get it right
- Enter the total band, not the ± value: ±0.02 mm is a 0.04 mm band.
What this means: Measured size
- What it is
- The size measured on the part with the gauge, corrected for any known gauge error.
- Units and format
- Enter a number in mm: millimetres. The form accepts -100000 mm to 100000 mm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 25.012 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Enter millimetres, the same unit as the nominal.
Uncertainty contributors
What this means: Calibration uncertainty of the gauge (expanded,…
- What it is
- The expanded uncertainty of the gauge from its calibration certificate, stated at a coverage factor of 2; the tool halves it to a standard uncertainty.
- Units and format
- Enter a number in µm: micrometres (µm, microns). The form accepts 0 µm to 1000 µm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 2 µm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Enter micrometres. If the certificate uses another coverage factor, convert to the k = 2 value first.
What this means: Resolution of the gauge
- What it is
- The smallest step the gauge display or scale can show; the tool turns it into a standard uncertainty.
- Units and format
- Enter a number in µm: micrometres (µm, microns). The form accepts 0 µm to 1000 µm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 1 µm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Enter micrometres: a 0.001 mm resolution is 1 µm.
What this means: Repeatability (standard deviation of repeat rea…
- What it is
- The standard deviation of repeated readings of the same feature with the same gauge, from a repeatability study.
- Units and format
- Enter a number in µm: micrometres (µm, microns). The form accepts 0 µm to 1000 µm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 1.5 µm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Use the standard deviation, not the range of the readings, in micrometres.
What this means: Abbe offset
- What it is
- The Abbe offset: the distance between the line along which the size is measured and the line of the gauge’s scale.
- Units and format
- Enter a number in mm: millimetres. The form accepts 0 mm to 2000 mm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 30 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- The Abbe error rises in direct proportion to the offset.
- Get it right
- For a micrometer it is near zero; for a caliper it is the distance from the jaws’ contact points to the beam scale.
What this means: Tilt of the moving element
- What it is
- The possible tilt of the moving jaw or slide during measurement, which with the Abbe offset causes a size error.
- Units and format
- Enter a number in arcsec: arcsec. The form accepts 0 arcsec to 3600 arcsec. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 20 arcsec as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Enter arcseconds (1° = 3600 arcsec), from the gauge specification or an estimate of slide play.
What this means: Length measured
- What it is
- The length over which thermal expansion acts in the temperature contributor, normally the measured length of the feature.
- Units and format
- Enter a number in mm: millimetres. The form accepts 0.000001 mm to 100000 mm. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 500 mm as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- The temperature uncertainty rises in direct proportion to this length.
- Get it right
- Use the length of the feature being measured; it is used only for the temperature term.
What this means: Part expansion coefficient
- What it is
- The coefficient of thermal expansion of the part material, from the material data sheet.
- Units and format
- Enter a number in µm/m·K: micrometres per metre per kelvin. The form accepts 0 µm/m·K to 50 µm/m·K. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 11.5 µm/m·K as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Enter micrometres per metre per kelvin, which equals 10⁻⁶ per kelvin.
What this means: Uncertainty of the temperature difference
- What it is
- The uncertainty in the temperature difference between the part and the gauge at the time of measurement.
- Units and format
- Enter a number in K: kelvin (a difference of 1 K equals 1 °C). The form accepts 0 K to 50 K. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 0.5 K as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- This is the doubt in the difference between part and gauge, not the room temperature or its distance from 20 °C.
What this means: Coverage factor
- What it is
- The coverage factor that multiplies the combined standard uncertainty to give the expanded uncertainty used for guard banding.
- Units and format
- Enter a plain number with no unit. The form accepts 1 to 4. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 2 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Effect on the result
- A larger factor widens the guard band and narrows the acceptance zone.
- Get it right
- Use the factor your quality system or the guidance on measurement uncertainty requires.
What this means: Least test uncertainty ratio
- What it is
- The smallest test uncertainty ratio you accept: the tolerance band divided by twice the expanded uncertainty.
- Units and format
- Enter a plain number with no unit. The form accepts 1 to 20. Required: the tool does not calculate until every input is valid.
- Example
- The form opens with 4 as a worked example. It is an illustration, not a recommended value: replace it with your own figure.
- Get it right
- Take it from your quality system or customer requirement; it is a ratio, not a percentage.
Method and formulas
- Standard uncertainties: calibration U ÷ 2, resolution ÷ √12, repeatability, Abbe and temperature errors ÷ √3; combined by root sum of squares; expanded = k × combined (as the measurement uncertainty budget tool).
- Test uncertainty ratio = tolerance ÷ 2U; acceptance zone = ±(tolerance ÷ 2 − U) about nominal.
- Accept inside the acceptance zone; reject outside the tolerance; otherwise indeterminate.
- Example: U = 5.9696 µm, so accept within ±19.03 µm; the part is 12 µm from nominal.
Limitations
- Symmetric tolerance and uncorrelated contributors.
- Simple guard band of one expanded uncertainty; your standard may use a different rule.
- Example values are illustrations. Replace them with values for your actual task.
Typical uses
- Decide whether a borderline part passes
- Set guard-banded acceptance limits on an inspection plan
- Show a customer how the decision was made
Who it is for
- Inspectors
- Quality engineers
- Metrologists
What you get
- Uncertainty budget, expanded uncertainty and test uncertainty ratio
- Guard-banded acceptance zone
- Accept, reject or indeterminate
Interpreting the result
- Indeterminate parts need a better measurement, not a coin toss: re-measure with a more capable gauge.
References
- Machinery’s Handbook, 29th edition (Industrial Press, 2012), printed pp. 709–735. Topic reference; no source tables or recommended settings are reproduced.
- ISO/IEC Guide 98-3 Uncertainty of measurement (GUM)
Learn the Concepts Behind This Tool
These BIZOMAT Library readings explain the method this tool uses, with worked examples.
Method explained here · BIZOMAT original book
Guard bands and the cost of errors
In Accept or Reject by the Numbers
What it covers: Set acceptance limits inside the tolerance: the one-uncertainty guard band in the BIZOMAT tool, cost-based limits from the costs of each error, sensitivity to costs, and contract rules.
BIZOMAT tools are screening and decision-support utilities. Verify important financial, engineering, safety, legal or operational decisions using appropriate professional standards and your organisation’s approved data.